Abstract

Molybdenum concentrations in eleven USGS geochemical reference materials AGV‐1, BCR‐1, BHVO‐1, BIR‐1, DNC‐1, DTS‐1, G‐2, GSP‐1, MAG‐1, PCC‐1 and W‐2 were measured by isotope dilution thermal ionisation mass spectrometry (ID‐TIMS). In every case but one, the concentrations determined in this study were significantly lower than the current consensus values. Molybdenum concentrations determined by ID‐TIMS are inherently more accurate and precisions may be up to an order of magnitude higher than those measured by other analytical techniques.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.