Abstract

Molecular absorption with electrothermal volatilization in a graphite tube. Part 3. A study of the determination of fluoride traces by A1F, GaF, InF and T1F molecular absorption. The use of the molecular absorption of thermally stable molecules such as AlF, GaF, InF and TlF, generated in a normal graphite tube, is assessed for determinations of fluoride. Optimal experimental conditions are discussed. The presence of Na + ions increases the signals. The influence of Na + ions and of other substances is discussed. The sensitivities (0.01 absorbance) are 0.4 ng F - for AlF, 0.8 ng F - for GaF, and 1.1 ng F - for InF.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.