Abstract

Different degrees of glycosaminoglycan sulfation result in their different charge densities. The charge density differences impact their migration behavior in polyacrylamide gel electrophoresis and size exclusion chromatography, two of the most common methods for determining relative molecular masses of polysaccharides. In this study, we investigated the feasibility of using commercially available heparin oligosaccharides as calibrants for measuring the relative molecular masses of intermediates in a bioengineered heparin process that have different levels of sulfation. A size exclusion chromatography method was established that eliminates this charge density effect and allows the determination of relative molecular mass using a single calibration curve with heparin oligosaccharides calibrants. This is accomplished by overcoming the electrostatic interaction between the glycosaminoglycans and size exclusion chromatography stationary phase using high ionic strength mobile phase.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.