Abstract

A reliable method is described for analyzing ultrahigh molar mass polyethylene (UHMW-PE) with a weight-average molar mass of 5700 kg/mol and molar mass fractions up to 50,000 kg/mol by size exclusion chromatography (SEC) without molecular degradation as was proved by comparison of on-line and off-line viscometry. The use of low-bleed SEC columns (PLgel 20 µm Mixed-A LS), especially designed for on-line light-scattering detectors, allowed SEC to be coupled to a low-angle laser light-scattering detector without a filter ahead of the light-scattering cell. With this procedure, molded specimens of UHMW-PE, which are used for medical implants, such as hip endoprostheses, containing various amounts of f -tocopherol as a biocompatible stabilizer, were investigated. Furthermore, the effect of sterilization of the samples by n -irradiation ( 60 Co, 1.33 MeV, absorbed dose: 25 kGy) was studied with respect to the content of insoluble cross-linked material and the molar mass distribution of the soluble fraction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.