Abstract

α,β-Poly(asparthylhydrazide) (PAHy) is a new synthetic polymer that exhibits interesting properties and is a candidate for biomedical applications. In this article the characterization of PAHy polymer by multi-angle laser light scattering (MALS) and single-capillary viscometer (SCV) detectors on-line to a size exclusion chromatography (SEC) system is reported. The SEC–MALS–SCV system furnishes exhaustive and consistent characterization of the PAHy polymer. Further, it is possible to characterize the PAHy polymer through conventional SEC and universal calibration. The universal calibration method gives intrinsic viscosity and dispersity very close to those measured by the absolute detectors; instead the weight-average molar mass is approximately 8% lower. This finding means that the chromatographic separation of PAHy polymer is fundamentally based on its hydrodynamic volume. By the SEC–MALS–SCV system the constants of the Mark–Houwink–Sakurada equation were also estimated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.