Abstract

Cobalt-chrome modular distal ulnar head replacement arthroplasty is a surgical option to restore stability to the distal radioulnar joint rendered unstable by hemi-resection arthroplasty or a total resection arthroplasty. However, the revision of dislocated modular cobalt-chrome ulnar head implants may pose an important intraoperative challenge. The Morse-taper disassembly force of modular ulnar head implants is not available in the current published literature. We present a case in which tremendous difficulty was encountered while revising a dislocated modular cobalt-chrome distal ulnar head implant. The mean Morse-taper disassembly force of the retrieved modular cobalt-chrome implant was 2958 N +/- 1272. At nearly 4.5 times the average body weight, the modular ulnar head Morse-taper disassembly strength presented a formidable force to overcome intraoperatively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.