Abstract

Modular necks were introduced in total hip arthroplasty (THA) to improve restoration of hip-geometry and reduce dislocation-rates. This presumed advantage was evaluated retrospectively for patients with arthritis in otherwise anatomically normal hips. Restoration of hip-geometry was assessed on preoperative and postoperative calibrated radiographs in 95 consecutive primary THAs with a modular neck design and compared with 95 match controlled THAs with a similar monoblock stem. No significant differences were seen in restoration of body moment arm, leg length and cupangle. Offset restoration revealed a borderline significant difference (P=0.48) with higher values for the monoblock stem. In both groups 4 dislocations within one year were encountered. In this study modular necks did not reveal a clear benefit in restoring hip geometry and dislocation rate after straightforward THA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.