Abstract

The total hip prosthesis has evolved from the initial monoblock system developed by Charnley. Modularity of both the femoral and acetabular components, which allows the surgeon to assemble the prostheses just before or during the course of implantation, has now become standard. Modularity increases the number of implant parameter variables, allowing semicustom devices to be constructed. Component interfaces, however, are subject to significant stresses, micromotion manifesting as fretting, may produce wear debris, and at times may experience macromotion leading to catastrophic failure. We report 3 cases of stem-sleeve junction failure in otherwise uncomplicated primary total hip arthroplasties. Two of the failures occurred in the S-ROM (Joint Medical Products, Stamford, Conn) femoral stem design and the other in an Apex stem (Omni Life Science, Raynham, Mass). To our knowledge, there have been no previous reports of clinical failures at femoral stem-sleeve junction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.