Abstract

A modular ceramic femoral head was selected for use as the bearing on a novel carbon fibre reinforced plastic (CFRP) total hip replacement femoral stem. Mechanical testing was required to validate this novel combination of materials for load bearing implant application. The Ceramic Head-CFRP Stem Taper interface was tested under axial fatigue loads according to BS/ISO 7206-5, followed by static loading to failure. The assembly survived the fatigue loads and exceeded the required FDA post-fatigue strength by 50%. In parallel, the test was simulated in a Finite Element Analysis study, resulting in recommendations for the adhesive bonding technique used to attach the modular head to the CFRP stem to further improve its strength.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.