Abstract

An apparatus, hereinafter named the modified vacuum thermal stability (MVTS) test apparatus, that replaces the mercury manometric method used to determine the vacuum thermal stability of energetic materials is described. The volume calibrated MVTS system provides continuous pressure/time data for the duration of the test via a computerized data acquisition system. Total gas volume (reduced to STP) is computed from the pressure data. Subsequent gas chromatographic analysis of the evolved gaseous mixture further provides identification and individual volumes of the components. Water/solvent volume is obtained by difference (Vp — VCalibrated Gases GC).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.