Abstract

A recently introduced fluorescence polarization immunoassay (FPIA) for determination of cyclosporine A in serum and plasma is discussed with regard to its use for whole-blood samples, with and without hemolysis before the assay. The performance characteristics of the modified method are highly satisfactory (within-run CVs 2.06 to 5.50% and 1.99 to 3.39%, respectively; long-term between-run CVs under routine conditions 5.73 to 8.95%). The limit of detection is 30 micrograms/L. Results agree well with those obtained with the RIAs compared, but the modified FPIA is more convenient and faster.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.