Abstract

We report the design of a modified resectoscope loop for transurethral resection of the ejaculatory ducts. The loop is a modification of a standard 24 F resectoscope loop. Cutting dimensions are 2.5 mm in width and 3 mm in depth for the modified loop versus 6 mm in width and 5 mm in depth for the conventional loop. This modification allows precise resection of obstructed ejaculatory ducts with minimal trauma to the prostatic fossa, reducing the potential for troublesome bleeding.