Abstract

Recombinant hirudin (r-hirudin) is being used increasingly for therapeutic anticoagulation in patients with heparin-induced thrombocytopenia undergoing cardiovascular surgery. Although multiple laboratory methods are available for measuring r-hirudin, the ecarin clotting time (ECT) is the most commonly used for this purpose. Ecarin (extracted from snake venom) converts prothrombin to meizothrombin, which promotes clot formation. Direct thrombin inhibitors, like r-hirudin, bind meizothrombin and yield a linear, dose-dependent prolongation of ECT. Low levels of prothrombin and fibrinogen in plasma samples can lead to higher ECT, suggesting falsely elevated r-hirudin levels. A modified ECT assay with prothrombin and fibrinogen in excess was optimized using an orthogonal array method to eliminate the variations in patients’ plasma prothrombin and/or fibrinogen levels for accurate determinations of plasma r-hirudin levels. By using the modified ECT assay, falsely elevated r-hirudin levels can be avoided in patients undergoing cardiopulmonary bypass, thus providing reliable and accurate r-hirudin monitoring in this clinical setting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.