Abstract

In order to facilitate intra-operative use of magnetic resonance imaging (MRI) in neurosurgery an MRIcompatible headholder was developed and adapted to a modified MR-couch simultaneously serving as tabletop for the operating table. To allow shock-free transport into the scanner the wheels of the operating table were replaced by an air cushion mechanism. In 75 procedures the system proved to be reliable and safe. Image quality was not impaired by the fixation device. With growing routine the transfer became straightforward/ requiring approximately 70 min. Intra-operative MRI is thus made possible with minimal changes to the standard surgical environment. Its benefit however/ still remains to be critically investigated. [Neural Res 1998; 20: 658–661]

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.