Abstract
BackgroundMethotrexate is an important component in many chemotherapy protocols. The blood concentration of Methotrexate is used to determine the regimen of folinic acid. However, the lower limit of Siemens assay kit is 0.30 μmol/L in China. This study extended the limit from 0.3 to 0.05 μmol/L and reduced the test cost by optimizing the parameters of Enzyme Multiplied Immunoassay Technique assay.MethodsParameters of Enzyme Multiplied Immunoassay Technique assay were modified to decrease the volume of reagents A and B. Then a standard curve with a new custom set of calibrators was prepared to detect low concentration. Intra-day and inter-day imprecision were assessed by control material and samples. The linearity of the modified assay was verified by analyzing a range of quality controls with known concentration from 0.05 to 1.00 μmol/L. At last, the same samples were tested by modified Enzyme Multiplied Immunoassay Technique assay and Liquid Chromatography-tandem Mass Spectrometry assay respectively. A simple linear regression was performed to verify the validity of the modified Enzyme Multiplied Immunoassay Technique assay.ResultsIntra-day and inter-day imprecision show good reproducibility at all levels (0.05, 0.12, 0.43, 0.82 μmol/L). The linearity equation of modified assay was y = 0.9913x + 0.0046, in which y was the mean of measured concentration and x was the target concentration (R2 = 0.9994). In the range of 0.05–10.00 μmol/L, correlation between the Modified assay and Liquid Chromatography-tandem Mass Spectrometry assay was significant (r = 0.9968). In the range of 0.30–10.00 μmol/L, the correlation between modified and commercial assays was significant (r = 0.9987) as well.ConclusionsThe modified assay enhanced the sensitivity of Siemens VIVA-E to 0.05 μmol/L. In addition, the test number of a reagent Kit increased from 140 to 210. This means the cost of detection was reduced about 30%.
Highlights
Methotrexate is an important component in many chemotherapy protocols
Folinic acid rescue is continued until MTX concentration fall below
The lower limit of quantitation decreased from 0.30 μmol/L to 0.05 μmol/L, and the cost of detection was reduced about 30%
Summary
The blood concentration of Methotrexate is used to determine the regimen of folinic acid. This study extended the limit from 0.3 to 0.05 μmol/L and reduced the test cost by optimizing the parameters of Enzyme Multiplied Immunoassay Technique assay. When high dose of MTX (HDMTX) is administered intravenously and followed by rescue with folinic acid, it is crucial to monitor the blood concentration of MTX at the fixed time points to determine the optimal dose and regimen of folinic acid. The lower limit of this assay is just 0.30 μmol/L, which is insufficient to meet the diagnostic criterion to terminate folinic acid rescue. The lower limit of quantitation decreased from 0.30 μmol/L to 0.05 μmol/L, and the cost of detection was reduced about 30%
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.