Abstract

The purpose of this work was to evaluate electron beam CT (EBCT) for the noninvasive assessment of modified Blalock-Taussig (BT) shunt patency in patients with pulmonary atresia. Five infants and children with pulmonary atresia and modified BT shunts underwent contrast-enhanced EBCT. Modified BT shunts from the subclavian artery to the pulmonary artery were performed to improve the pulmonary blood flow. Electrocardiogram (ECG)-triggered EBCT was obtained with a 100 ms exposure, 3 mm section thickness, and 2 mm table feed after intravenous administration of contrast material. Three-dimensional (3D) or maximum intensity projection (MIP) EBCT images were compared with conventional angiography. The visibility of modified BT shunts was graded and recorded with use of a four-point scale. Satisfactory visualization was achieved in both 3D and MIP EBCT images to evaluate modified BT shunt patency. Contrast-enhanced 3D or MIP EBCT imaging with ECG trigger may be used as an effective substitute to evaluate modified BT shunts with low radiation dose exposure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.