Abstract

Radioactive gold (198Au) grains are used extensively as interstitial sources of radiation in the treatment of malignant disease (Sinclair, 1952). Usually the grains are implanted by means of a special gun. The original instrument was designed at the Royal Marsden Hospital (Hodt, Sinclair and Smithers 1952) and since then similar models have been manufactured by W. H. Bailey & Son Ltd. (M.S.A.) The grains are contained in aluminium magazines which fit into the barrel of the implantation gun. The original method of loading the gun consisted of the following stages: (a) The aluminium caps of the magazine were removed with a screwdriver. (b) In order to prevent the radioactive gold grains from falling out of the magazine, its ends were dabbed with sterile petroleum jelly. (c) The magazine was inserted into the barrel of the gun, which in turn was screwed onto the shaft. (d) The gun could now be primed ready for use. Exposure to radiation during this procedure was chiefly to the hands, fingers and eyes, and wa...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.