Abstract

In order to improve blood compatibility of polyethylene (PE) film, the Pluronics F127 additives in the PE film were then crosslinked to be stably entrapped in the PE matrix. The crosslinking was done by free radicals produced from the decomposition of dicumyl peroxide (DCP) in the film through heating (120 °C). Surface properties of the Pluronics F127 additive-containing PE films were investigated by Fourier transform infrared spectroscopy (FTIR), electron spectroscopy for chemical analysis (ESCA) and water contact angle (WCA) measurements. The blood compatibility of the Pluronics F127 additive-containing films was evaluated by platelet-rich plasma and blood-cell adhesion tests, respectively. And the results were observed by scanning electron microscopy. The blood compatibility of the prepared Pluronics F127 additive-containing film is better than that of blank PE film. These results suggest that the blood compatibility of Pluronics F127 additive-containing films make them suitable biomaterials for some applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.