Abstract

AbstractIn order to have available a specimen holder suited to measure the beam current as is often required in quantitative electron probe X‐ray microanalysis, the rod of a low background beryllium specimen holder of a transmission electron microscope was modified. The tip was electrically insulated from the mass of the microscope and connected electrically to the central contact of a BNC connector mounted on the specimen holder handle. With this modified specimen holder the current absorbed by the specimen and/or the specimen holder could be measured easily and accurately. The modified specimen holder has been used to measure the beam current stability of an analytical electron microscope under various conditions. Data were obtained for tungsten as well as lanthanum hexaboride cathodes. Small changes to other types of specimen tips made it possible to exchange these for the low background tip.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.