Abstract
A simple and inexpensive approach to convert the electrospray nebulizer of a commercial liquid chromatography/mass spectrometry (LC/MS) system (HP 1100) to accommodate lower flow rates has been proposed and evaluated. This modification consists of simply replacing the nebulizer needle by a commercially available stainless steel needle with a smaller internal diameter. Experiments were conducted in order to optimize operational parameters. Using two different internal diameter needle sizes, flow rates ranging from 1 to 250 microL/min could be accommodated. The modification presented allows an extension of the range of compatible flow rates without major modification of the standard design of the interface.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.