PurposeThe purpose of this study is to establish Loratadine [LRD] quantification in purified and capsule formulations using a precise and specific Reversal Phase with a very high-performance liquid Chromatographic [RP-HPLC] technique. The approach was evaluated in agreement with the principles of the International Conference on Harmonization [ICH]. Arcus EP-C18 Ion Pac column, 5 m, 4.6 mm, 250 mm, mobile phase Methanol: Acetonitrile (60:40) v/v. Dibasic potassium phosphate buffer, pH 7.2, flow rate 1.0 ml/min.Design/methodology/approachThe HPLC system used a 340 nm UV detector for testing. A 10-min run time was used for the analysis. At concentrations ranging from 2 to 10 g/ml, the technique was linear (R2 = 0.9998), exact (intra-day and inter-day relative standard deviation [RSD] values 1.0%), accurate (range recovery = 96%–102%), exclusive and strong.FindingsThe detecting and quantitation limits were 0.92 g/ml and 2.15 g/ml, respectively.Originality/valueThe findings demonstrated that the proposed method could accurately determine LRD in bulk and pill dose formats quickly and accurately.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.