Abstract

The electrical properties of cervical squamous epithelium have been modelled in the frequency range 100 Hz to 10 MHz. The hierarchical modelling process comprises a cellular level stage, which includes detailed models of cells typical of different depths within the epithelium and a tissue model, which utilizes electrical properties obtained from the cellular models. The fit between the modelled and measured impedance spectra and the distribution of current with depth depends on the macroscopic model structure. Both the properties of the basement membrane and the presence of a surface mucus layer are shown to have a significant effect. The best fit with measured data is obtained when a 10 μm thick, high-conductivity surface layer is included in the tissue model.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.