Abstract

Moiré fringe method in X-ray grating interferometry is characterized by its advantage to obtain multi-contrast data through single-frame imaging. However, how the visibility of the Moiré fringe is influenced by the system parameters, such as the misalignment angle, still lacks investigation, although it closely relates to the signal-to-noise ratio of the image data. In this paper, a simplified model of the Moiré fringe visibility is presented, which reveals that the Moiré fringe visibility can be expressed as the product of a misalignment-angle-related "sinc" function and a relatively independent factor. The following experimental results further suggest that the crosstalk between the detector pixels in the direction perpendicular to the Moiré fringe orientation is another main cause for the visibility reduction of the Moiré fringes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.