Abstract

A model was developed to examine variables that influence signal-to-noise ratio (SNR) at the tympanic membrane (TM) when using a hearing aid (HA) and frequency modulated (FM) system. The model was used to explore how HA coupling influences SNR. To generate the model, HA output was measured in a coupler. Known coupler to real-ear transformations and known values for vent (gain) loss as a function of coupling were also used. The model was verified by measuring sound pressure level (SPL) at the TM in 6 ears. The model predicts similar overall SNRs at the TM regardless of coupling method when HA and FM microphones are active. The primary difference in SNR is in the low frequencies and depends on the amount of low frequency insertion gain and the noise levels at the HA and FM microphones. A model was developed to explore how complex variables contribute to SNR at the TM. One variable, HA coupling, is predicted to have only a minimal effect on SNR at the TM when there is HA gain. Further studies will be needed to assess the real-world effectiveness of an FM system coupled to an open- versus closed-fit HA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.