Abstract
Coriolis mass flowmeters are used for many applications, including as transfer standards for proficiency testing and liquified natural gas (LNG) custody transfer. We developed a model to explain the temperature dependence of a Coriolis meter down to cryogenic temperatures. As a first step, we tested our model over the narrow temperature range of 285 K to 318 K in this work. The temperature dependence predicted by the model agrees with experimental data within ± 0.08 %; the model uncertainty is 0.16 % (95 % confidence level) over the temperature range of this work.Here, basic concepts of Coriolis flowmeters will be presented, and correction coefficients will be proposed that are valid down to 5 K based on literature values of material properties.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.