Abstract

Performing photothermal surface displacement (PTD) technique at λ=10.6 μm in a vacuum chamber, the measured signal is obtained by the change in probe beam deflection at the bulge of the reflecting heated sample surface. The influence of the accuracy-limiting quantities such as angle of incidence of the probe beam as well as its finite lateral extension, and the detector geometry and the temperature-dependent reflection at the sample surface has been investigated by a numerically performed simulation in order to optimize the measuring procedure. The maximum resolution power is achieved for probe beam diameters small compared to the heating beam diameter. An unambiguous measurement may be carried out satisfying the following condition between the quad-cell slit width dslit, the probe beam diameter 2rPB, and the heating beam diameter 2rHB: 2rHB≥2rPB≫dslit keeping the probe beam angle of incidence αPB as small as possible. Moreover, it has been shown that the contribution of the temperature influenced reflection on the change of the probe beam intensity can be also neglected at λ=10.6 μm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.