Abstract

This study developed a finite element (FE) model of simulated locking plate fixation to examine the strain response following supracondylar femoral plate attachment and under compressive loading. An implicit FE model of a synthetic femur with a distal fracture gap stabilized with a lateral plate was evaluated following attachment and 500 N loading, considering locking and non-locking proximal screws configurations. Screw pre-tension values of 60 N for both distal and proximal non-locking screws yielded good agreement with plate experimental strain data in attached (unloaded) and loaded conditions. The results highlight the importance of pre-tensioning in modeling plate attachment using non-locking screws.

Full Text
Published version (Free)