Abstract

We model the deformation of a thromboembolus lodged in a cerebral artery under the application of aspiration pressure as it would be provided by an aspiration catheter during a mechanical thrombectomy procedure. The system considered consists of (i) a clot modeled as a viscoelastic solid; (ii) an artery modeled as a hyperelastic solid; and (iii) a viscoelastic cohesive interface between the clot and the artery. For the chosen system and geometry, we show that the application of aspiration pressure results in the impingement of the thrombus against the inner arterial wall near the aspiration location. Conditions leading to interfacial failure are nucleated at the distal end of the clot and, depending on the details of the loading conditions, propagate toward the proximal end. The results provide useful information in identifying the circumstances that play a decisive role for clot removal by aspiration alone.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.