Abstract

In a cross capacitor with guard electrodes, the equivalent electrical length is sensitive to any variation in the cross section of the main electrodes unless measures are taken to compensate for this. The problem arises particularly in the case of a moveable guard electrode, which is conventionally fitted with a nosepiece of such geometry as to compensate for variations in the geometry of the main electrodes from one guard position to another. In the new NIM cross capacitor an alternative approach, in which a fraction of the measuring voltage is applied to an isolated electrode in the face of the guard electrode, has been investigated. Results indicate that this active nosepiece performs better than an equivalent geometrical compensating nosepiece.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.