Abstract

We present a method to calibrate Raman measurements of reactive systems without relying on thermodynamic models. Raman spectroscopy is able to detect multiple species even in reactive mixtures. Classical calibration methods require the knowledge of concentrations of all species as input. Typically, concentrations of intermediates cannot be prepared and fixed independently and are therefore calculated using thermodynamic models. The quality of the results thus depends on the accuracy of the thermodynamic model. In this paper, we present a method based only on stoichiometric balances and electroneutrality. By avoiding the use of thermodynamic models in the calibration step, the risk of overfitting spectroscopic data to a thermodynamic model is avoided. The presented method is demonstrated for the reactive system monoethanolamine, water, and CO2 and is validated by a thermodynamic model taken from the literature. While the model-free calibration is demonstrated for Raman spectroscopy in this work, the approach is generic should thus be applicable to most spectroscopic techniques.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.