Abstract

This study investigated the efficacy of a combined high-speed, biplane radiography and model-based tracking technique to study hip joint kinematics and arthrokinematics. Comparing model-based tracking to the gold standard of radiostereometric analysis using implanted metal beads, joint translation was measured with a bias of 0.2 mm and a precision of 0.3 mm, whereas joint rotation was measured with a bias of 0.2° and a precision of 0.8°. A novel measure of hip arthrokinematics characterizing the region of closest contact in the anterosuperior acetabulum was measured with a bias of 0.9% and a precision of 2.5%. Model-based tracking of the hip thus provides the opportunity to noninvasively study hip pathologic conditions such as osteoarthritis and femoroacetabular impingement with great accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.