Abstract

Abstract This paper presents an optimization-based control scheme for a novel left ventricular assist device. This cardiac support system consists of two pumps connected in series and a buffer reservoir in between them. This novel concept for relieving the left ventricle is designed to allow the pre- and afterload on the heart to be explicitly adjusted and independently of each other. The first pump controls the preload of the heart with a known physiological controller. With an iterative model-based optimization, the afterload is minimized by the second pump while complying with all constraints. The proof of the control concept and a comparison to the classical single pump left-sided cardiac support is performed on a hardware-in-the-loop test bench. The results show that with the new left ventricular assist device concept, the afterload can be reduced compared to the classical cardiac support.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.