Abstract
The authors advise on an exceptional complication in a totally implantable venous access and give recommendations on how to avoid this problem. In an obese person, one week after the implantation of a totally implantable venous access, a chest X-ray showed that the catheter was too short. In the operative room, at the time of the second intervention, the radioscopy did not find this anomaly. The subcutaneous injection port was positioned too low. When the patient was in a supine position, the subcutaneous injection port was in the correct position but when she was in an upright position, the brest descended and pulled down the subcutaneous injection port and the catheter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.