Abstract
1. Understand the various origins of MRA contrast mechanisms that do not require a contrast agent. 2. Understand the issues related to NCE-MRA imaging including design, acquisition and processing. 3. Understand the benefits, pitfalls, and future potentials of these approaches. My research is sponsored by GE Healthcare.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.