Abstract

The multiplicative Rayleigh noise model has been used for maximum likelihood (ML) motion estimation in ultrasound imaging (UI). In this work, we introduce new robust similarity measures that take into account the deviations from the Rayleigh statistics resulting, for example, from multiple scatterings or acquisition artefacts. Specifically, the $t$ -distribution is used for modelling the radio-frequency (RF) signals and the Nakagami-Gamma (NG) model is used for the echo amplitudes. Experiments using in vivo images of the carotid artery show an improvement in motion estimation accuracy in comparison with the similarity measure based on the classical Rayleigh model.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.