Abstract

In the phenomenon of mixed-mode oscillations, transitions between large-amplitude and small-amplitude oscillations may lead to anomalous jitter in the probe of a tapping mode atomic force microscope (TM-AFM) during the scanning process, thereby affecting the accuracy and clarity of the topographical images of the tested sample's surface. This work delves deeply into various mixed-mode oscillations and the corresponding formation mechanisms in TM-AFM under low-frequency resonant excitation. Through a detailed analysis of bifurcation sets of the fast subsystem, we found that the system's mixed-mode oscillations encompass the typical two coexisting branches and the novel three coexisting branches of equilibrium point attractors. In the stable case, a certain transition pattern in phase trajectory can be observed involving two jumps and four jumps, switching between quiescent and spiking states. In the bi-stable case, the trajectory undergoes distinct transitions decided by whether to pass through or crossover the middle branch of attractors when bifurcation occurs. By applying basin of attraction and fast-slow analysis methods, we unfold the dynamic mechanism of mixed-mode oscillations with distinct switching patterns. Our research contributes to a better understanding of complex oscillations of TM-AFM and provides valuable insights for improving image quality and measurement precision while mitigating detrimental oscillations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.