Abstract

A novel dual-mechanism ionization technique for LC/MS/MS has been observed, characterized and applied to the quantitation of a tertiary amine-containing drug compound in dog plasma. This mixed-mechanism ionization approach can improve the sensitivity of the pneumatically assisted electrospray experiment. Under conditions of higher than normal chromatographic flow and lower electrospray voltage, approximately a 4-fold increase in sensitivity was realized. A detection limit of 16 pg (45 fmol) on-column, and inter-day imprecision and inaccuracy of <11 and <15%, respectively, were obtained. A trade-off in concentration sensitivity in favor of ease of sample preparation was made to increase sample throughput. Although results strongly suggest that mixed-mechanism ionization is in operation, and that pneumatically assisted electrospray is a partial contributor to the overall ionization process, the exact nature of the second mechanism of ionization is unclear at this time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.