Abstract

Optimal translocation patch width for functional mitral regurgitation (FMR) treatment was evaluated in an air-filled ex vivo system. FMR was created in 19 isolated swine hearts by annular dilation and papillary muscle displacement. Frustum-shaped pericardial patches of varying widths (Group 1 = 0.5cm; Group 2 = 1.0cm; Group 3 = 1.5cm) were implanted and imaged via a 3D-structured light scanner. Median leaflet coaptation decreased (P < 0.001) from 5.5 ± 2.0mm at baseline to 2.4 ± 1.3mm following FMR creation. Translocation repair increased coaptation length over FMR levels by 2.2mm in Group 1 (P < 0.001), 4.6mm in Group 2 (P < 0.001), and 4.7mm in Group 3 (P < 0.001). After repair, no significant differences were found between groups for annular height, circularity index, tenting height, tenting area, and non-coapting surface area. The supranormal coaptation and minimal valve geometric changes support using a 1.0- or 1.5-cm translocation patch for FMR treatment. Implantation of a 1.0-cm or 1.5-cm circumferential pericardial patch (mitral valve translocation) increases leaflet coaptation length without significantly altering valve geometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.