Cardiac output (CO) assessment in neonates is commonly done by echocardiography. It is unclear which is the best site to measure the left ventricular (LV) outflow tract for CO assessment (the aortic valve [AV] aortic sinus [AS] or the sinotubular junction [STJ]). In the normal heart, the blood flow entering the LV equals the blood ejected from it. Therefore, measuring the blood flow into the LV through the mitral valve (MV) is an alternative way to measure CO. In stable preterm infants the MV CO was compared with the right ventricular (RV) CO and the three ways to measure LV CO, in 30 stable preterm neonates. Interobserver variability for MV CO was established. In the 30 neonates studied, MV CO was best correlated and had a minimal bias to the RV CO and LV CO measured at the STJ. Left ventricular CO measured at the AV and AS had significant bias relative to RV CO and MV CO. MV CO inter-observer variability was similar to other echocardiographic CO assessment methods. MV CO may be used as an alternative way to assess CO. The STJ may be the optimal site to measure LV outflow tract.