Abstract

In 2022, the Food and Drug Administration approved a second mitral transcatheter edge-to-edge repair device for the treatment of primary mitral regurgitation (PASCAL Precision Transcatheter Valve Repair System, Edwards Lifesciences, Irvine, CA). The PASCAL Precision system consists of a guide sheath, implant system, and accessories. The implant system consists of a steerable catheter, an implant catheter, and the implant (PASCAL or PASCAL Ace). The guide sheath and steerable catheter move and flex independently from each other and are not keyed, allowing for freedom of rotation in three dimensions. This manuscript provides an overview of the PASCAL Precision system and describes the basic and advanced steering maneuvers to facilitate effective and safe mitral transcatheter edge-to-edge repair.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.