Abstract
X-ray diagnostics in today's high-energy density environments must contend with intense and energetic X-ray background levels. In this work, we address the issues of X-ray fluorescence and scattering in reflection-geometry X-ray crystal spectrometers. In this geometry, the detector can capture not only a dispersed X-ray spectrum but also fluorescence and/or scattered X-rays from the diffracting crystal and crystal mounts. Studies to optimally reduce these sources of spectral contamination have been performed using the HENEX spectrometer. Variables that mitigate such unwanted background include filtration, collimation and judicious selection of crystal and detector materials.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.