Abstract

COVID-19 is a novel virus spread via airborne particles. Given the inherent risk to the anesthesia provider, intubation and airway management guidelines have been recently established. Various studies have been published advocating and detailing the results of different intubation devices designed to decrease the number of airborne particles. Currently, little literature exists regarding devices designed to mitigate the spread of COVID-19 airborne particles during extubation. The purpose of this prospective in situ simulated manikin study was to measure the effectiveness of an aerosolized containment device during passive (deep) and forced (simulated coughing) extubation. Airborne particles were measured at the 0.3, 0.5, 1, 2, 5, 10-micron level. Statistically significant decreases were seen with the use of a barrier device during both passive and forced extubation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.