Abstract

Due to the low Young's Modulus of porous silicon (PS), Si/PS composite membranes - where the silicon membrane is converted into PS to a certain depth - deform more than silicon membranes and hence MEMS pressure sensors with composite membranes have higher sensitivity. But the Si/PS composite membranes exhibit a smaller range of linear response with applied pressure than silicon membranes with the linear range being less for Si/microPS as compared to Si/macroPS composite membranes. In addition, while the composite membrane deformation saturates at high pressures like silicon membranes, the deformation is irreversible unlike that seen with silicon membranes within reasonable limits. With the possibility that the irreversible deformation could be due to stiction force between the collapsed pore walls at high pressure, we investigate the effect of formation of self-assembled monolayer (SAM) antistiction coating on the performance of Si/PS composite membranes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.