Abstract

High-field MRI has the challenge of inhomogeneous B1(+) and consequently an inhomogeneous flip angle distribution. This causes spatially dependent contrast and makes clinical diagnosis difficult. Under the small flip angle approximation and using nonlinear spatial encoding magnetic fields (SEMs), we propose a method to remap the B1(+) map into a lower dimension coordinate system. Combining with RF shimming method, a simple pulse sequence design using nonlinear SEMs can achieve a homogenous flip angle distribution efficiently. Using simulations, we demonstrate that combining RF shimming and spatially selective RF excitation using generalized SEMs (SAGS) using linear and quadratic SEMs in a multi-spoke k-space trajectory can mitigate the B1(+) inhomogeneity at 7T efficiently without using parallel RF transmission.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.