Abstract
We present two cases of unrecognized endotracheal tube misplacements in out-of-hospital cardiopulmonary resuscitation recognized by the analysis of transthoracic impedance. In Case 1, ventilation-induced changes in transthoracic impedance disappeared after an intubation attempt corresponding to oesophageal intubation. This was clinically recognized after several minutes, the endotracheal tube was repositioned and alterations in transthoracic impedance resumed. In Case 2, the initial ventilation-induced signal change following endotracheal intubation weakened after a few minutes. At that time, the defibrillator gave vocal and visual feedback to the rescuers on ventilatory inactivity, a pharyngeal air leak was discovered simultaneously and the tube was found to be dislodged. Continuous monitoring of transthoracic impedance provided by the defibrillator during cardiopulmonary resuscitation may contribute to the early detection of an initially misplaced or later dislodged endotracheal tube.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.