Abstract

( A new standardized method for determination of ethylene oxide.) This paper describes a new standardized method for the chemical transformation of 2-chloroethanol into ethylene oxide. The reaction is developed in a miniatured glass stopped vial. The recovery is about 60%. The limit of detection was found to be 0.020 μ mol l −1. Ethylene oxide was determined by a classical head-space gas chromatographic method establishing a linearity of r = 0.999, good reproducibility and a relative standard deviation of 3.7%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.