Abstract

ABSTRACT Mirror velocity variations in a continuous mirror drive Michelson interferometer result in erroneous interferogram sampling. The effects of both systematic and random mirror fluctuations are evaluated. Four methods of evaluation provide a means to quantify the importance of mirror velocity errors in the Fourier spectrometer's performance as a spectroradiometer. The four methods are time interval signal analysis, sine wave test, optical chopper modulation, and examination of consecutive interferograms. Examination of consecutive interferograms includes the monitoring of a constant blackbody infrared radiation source as well as an interposed ammonia sample cell for wavenumber axis registration. Each method furnishes a known input to the interferometer modulation equation (i.e. 2uv=f) and permits an internal validation of the interferometer mirror velocity evaluation results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.