Abstract

This paper studies the optimum signal choice for the estimation of the aortic blood pressure via aorta radius, using a monostatic radar configuration. The method involves developing the Cramer-Rao lower bound (CRLB) for a simplified model. The CRLB for model parameters are compared with simulation results using a grid-based approach for estimation. The CRLBs are within the 99% confidence intervals for all chosen parameter values. The CRLBs show an optimal region within an ellipsoid centered at 1 GHz center frequency and 1.25 GHz bandwidth with axes of 0.5 GHz and 1GHz, respectively. Calculations show that emitted signal energy to received noise spectral density should exceed 1012 for a precision of approximately 0.1mm for a large range of model parameters. This implies a minimum average power of 0.4 µW. These values are based on optimistic assumptions. Reflections, improved propagation model, true receiver noise, and parameter ranges should be considered in a practical implementation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.