Abstract

We investigated the efficacy of combining highly cross-linked polyethylene with ceramic heads on further reduction in polyethylene wear compared with the combination with cobalt-chromium heads via PolyWare computer-assisted method. A prospective cohort study was performed on 102 cementless total hip arthroplasties using Longevity (Zimmer, Warsaw, Ind) highly cross-linked polyethylene liners. Either 26-mm zirconia heads or 26-mm cobalt-chromium heads were randomly used in 51 hips each. At a mean follow-up of 6.7 years, no significant differences were identified between the groups for total penetration rate and steady-state wear rate. Osteolysis was not observed in any hips in either group. In conclusion, no advantage was seen for the 26-mm zirconia head compared with the 26-mm cobalt-chromium head in this period.