Abstract
In this paper there are described three different algorithms of Minimum Detectable Activity (MDA) calculus, and its use in high resolution gamma spectrometry. In the first part, few introductive theoretical aspects related to the MDA are presented. Further, the theory was applied to real gamma rays spectrometry measurements and the results were compared with the activities reference values. Two different gamma spectrometry systems, both of them using High Purity Germanium (HPGe) detectors, but having different efficiencies, were used. Samples having different geometries and radionuclides content were measured. The measured samples were made by dissolving of some acids containing anthropogenic radionuclides in water, obtaining a density of 1g/cm3. Choosing this type of matrix was done because of its high homogeneity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.